标准摘要
[中文适用范围]: 本标准规定了绷膜型单桩骨内牙种植体及其预成修复瓤的疲劳试验方法.该方法zui适于比较不同设计、不同尺寸的骨内牙种植体。 虽然本标准模拟骨内牙种植体的主体和预成修复组件在“zui坏情况”条件下的功能载荷,但不适用于预测骨内牙种植体或修复体的体内性能,尤其不适用于多桩种植体修复的情况... [外文原描述]: ISO 14801:2016 specifies a method of dynamic testing of single post endosseous dental implants of the transmucosal type in combination with their premanufactured prosthetic components. It is most useful for comparing endosseous dental implants of different designs or sizes. This International Standard is not a test of the fundamental fatigue properties of the materials from which the endosseous implants and prosthetic components are made. This International Standard is not applicable to dental implants with endosseous lengths shorter than 8 mm nor to magnetic attachments. While ISO 14801:2016 simulates the functional loading of an endosseous dental implant under "worst case" conditions, it is not applicable for predicting the in vivo performance of an endosseous dental implant or dental prosthesis, particularly if multiple endosseous dental implants are used for a dental prosthesis.
英文名称Dentistry — Implants — Dynamic loading test for endosseous dental implants