标准摘要
[中文适用范围]: 本部分 ISO 16526 规定了使用 X 射线能谱(光谱法)对 X 射线管电压进行非侵入式测量的测试方法。它涵盖了 10 kV 至 500 kV 的电压范围。其目的是检查实际电压与 X 射线装置控制面板上指示值之间的对应关系。该程序仅旨在测量最大能量,而非完整的 X 射线能谱。该程序适用于罐式(tank type)和恒电位 X 射线装置。 [外文原描述]: ISO 16526-3:2011 specifies the test method for a non-invasive measurement of X-ray tube voltages using the energy spectrum of X-rays (spectrometric method). It covers the voltage range from 10 kV to 500 kV. The intention is to check the correspondence of the actual voltage with the indicated value on the control panel of the X-ray unit. It is intended to measure the maximum energy only and not the complete X-ray spectrum. The procedure is applicable for tank type and constant potential X-ray units.
英文名称Non-destructive testing — Measurement and evaluation of the X-ray tube voltage — Part 3: Spectrometric method